5 Signs Your Sampling System Is Not Performing

February 10 2017 Process Analytical

Process_stacks.jpg

An under-performing sampling system can cost an operator more than just time and money. A lousy system can be the cause of poor yield, slow response, bad decisions, and regulatory or contract violations. Is your process sampling system giving you reliable timely data, or is it giving you "alternative facts"?  How do you know when a system is not doing it's job?

Get 7 Tips For Improved   Analytical Sampling

Here are 5 signs you've got an under-performing analytical system.

 

1. Calibration errors

Excessive stabilization time, loss of low or high side calibration standard, slow recovery, poor peak shape and outright failure to get a calibration signal response can all be signs that your calibration system is not performing. A reactive calibration flowpath will result in lost peaks minimal peak area or peak tailing.

 

Reactive flowpaths show lost peaks or peak distortion

 

An inert flowpath will prevent signal loss, improve response and improve peak quality; making calibration faster and easier.

SilcoNert coated flowpaths are not reactive.  All peaks resolve with sharp peaks

2. Slow response

Tubing can be a huge source of sample loss and adsorption. How disruptive can reactive tubing be? Here's a comparative signal response graph plotting the response signal of a sulfur sample through 100ft of 1/8in tubing.1

Adsorptive surfaces result in significant delay in response

Active sites in the uncoated tube grab onto the sulfur sample, preventing it from reaching the analyzer for 90 minutes! Excessive delay in signal response will make process control and sulfur detection almost impossible.

Want to learn more about adsorption and desorption of active compounds and how to improve your inert sampling system? Read the complete study by Shell and O'Brien Analytical. Shell, O'Brien Analytical Adsorption Study

 

3. All is well, too well

If you've never seen a high sulfur reading, or results are "rock steady" and never changing, it's a sign that the sample system is adsorbing analyte or is saturated and desorbing at a consistent rate.

If you hear technicians state "we never have sulfur in our product", it could mean the sample system is adsorptive and not allowing your analyzer to detect sulfur. Poor response can make life easier sometimes, right up until regulatory or customer auditors show up and reveal the ugly truth.

 

4. Excessive maintenance

A reactive flowpath can be caused by corrosion. Poor system design or installation can result in moisture accumulation and corrosive particulate generation. Interaction with reactive iron oxide particles can grab onto a sample and prevent accurate signal response. If maintenance staff is constantly fixing system failures or replacing components due to corrosion, it could be a sign that your system could be causing adsorptive particle generation.

 

5. Variable response

If the system is showing no response and then an unexplained huge signal, the system may be adsorbing and then desorbing/releasing analyte. Seemingly random spikes can be a sign your system is holding active compounds then as the system becomes saturated, the surface may release or desorb analyte, causing sudden unexplained increases in response.1

Adsorbed sulfur can suddenly desorb and result in unexplained spikes in sulfur signal